Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
248User Manual 1173.9240.02 ─ 17
Parameters:
<TriggerLevel> For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST: -10 dBm
Example:
TRIG:LEV:IFP -30DBM
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:IQPower <TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed.
Parameters:
<TriggerLevel> Range: -130 dBm to 30 dBm
*RST: -20 dBm
Example:
TRIG:LEV:IQP -30DBM
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:RFPower <TriggerLevel>
This command defines the power level the RF input must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed. If defined, a reference level offset is also considered.
The input signal must be between 500 MHz and 8 GHz.
Parameters:
<TriggerLevel> For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST: -20 dBm
Example:
TRIG:LEV:RFP -30dBm
Manual operation: See "Trigger Level" on page 98
TRIGger[:SEQuence]:LEVel:AM:RELative <Level>
The command sets the level when AM-modulated signals are used as trigger source.
For triggering to be successful, the measurement time must cover at least 5 periods of
the audio signal.
Parameters:
<Level> Range: -100 to +100
*RST: 0 %
Default unit: %
Example:
TRIG:LEV:AM:REL -20 %
Sets the AM trigger threshold to -20 %
Configuring the Measurement