Remote Control - Commands
NRP2
216User Manual 1173.9140.02 ─ 07
Table 6-7: One-channel functions (<source_list> = (@n); n = 1,2,3,4 ).
-
Yields the measured average power of the sensor assigned to the PC.
:RELative
The value measured in the PC is divided by the reference value.
For high level commands, the calculation function is selected by the keyword in the
command header. The following calculation functions are available for calculating the
primary channel (PC) and the secondary channel (SC). The promary and secondary
channel are set by the parameter <source_list> in the parameter list.
Table 6-8: Two-channel functions ( <source_list> = (@n),(@m); n = 1,2,3,4; m = 1,2,3,4 )
DIFFerence
Yields the difference between the PC and SC. This means that two sensors must be
specified in <source_list>. This also applies to the following functions.
:DIFFerence:
RELative
value_reference
SCPC
The difference between the PC and SC is divided by the reference value.
:SUM
Yields the sum of the PC and SC. The <source_list> must therefore contain two sen-
sors. This also applies to the following functions.
:SUM:
RELative
The sum of the PC and SC is divided by the reference value.
:RATio
Yields the ratio of PC to SC.
:RATio:
RELative
The ratio of the PC and SC is divided by the reference value.
:SWR
If the forward power of a wave is measured in the PC and the reflected power in the
SC, the measurement command yields the standing wave ratio. The output unit is set to
"One" (UNIT:RAT O
).
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)
北京海洋兴业科技股份有限公司(证券代码:839145)