Remote Control - Commands
NRP2
205User Manual 1173.9140.02 ─ 07
*STB?
STatus Byte query reads out the contents of the status byte in decimal form.
Usage: Query only
*TRG
TRiGger generates a BUS-trigger event. If the sensor is in the WAIT_FOR_TRG state
and the source for the trigger is set to BUS (TRIG:
SOUR BUS), the sensor enters the
MEASURING state.
*TRG invalidates all current measuring results. A query of mea-
surement data following *TRG will thus always return the measured value determined
in response to *TRG.
Table 6-3: Error messages:
-211 Trigger ignored: No sensor is in the WAIT_FOR_TRG state.
-211 Trigger ignored: A sensor is in the WAIT_FOR_TRG state but the trigger source is not
BUS.
Usage: Event
*TST? <error code>
Self TeST query triggers the self test of the instrument and outputs an error code in
decimal form.
Parameters:
<error code> 0 | 1
0
no error
1
an error has occurred
Usage: Query only
*WAI
WAIt to continue prevents the execution of new commands until one of the following
criteria is met:
●
All previous commands have been executed.
●
A DCL command is received.
●
The device is switched off and on again (power-on-reset).
Usage: Event
Common Commands to IEEE 488.2