A.4 Programming Measurements
A.4.1 Measuring with high-level commands
The R&S
NRP2 is based on the instrument model defined in the SCPI 1999.0 stand-
ard. This model provides a number of logic blocks that can be configured via remote-
control commands. The R&S NRP2 is provided with high-level measurement com-
mands that automatically perform the most important settings and simplify measure-
ments, thus making it unnecessary to completely configure these blocks for each mea-
surement. Only the high-level commands are described, starting with the simplest
command, which is expanded in the following sections.
Note the following for the syntax of commands: There is a short and a long form for the
commands. The short form is normally used. However, commands are sometimes indi-
cated in their long form so that their meaning is easier to understand. The R&S NRP2
makes no distinction between upper case and lower case letters; they are only used to
distinguish between the short and long form.
A.4.1.1 The simplest measurement
The simplest way to obtain a measured value is to use the high-level measurement
command
MEAS?,
which configures sensor 1 in the ContAv mode, starts a measurement and always out-
puts a result that is stored in the output queue.
The following settings are performed:
INIT:CONT OFF
TRIG:SOUR IMM
TRIG:COUN 1
TRIG:DEL:AUTO ON
SENS:AVER:STAT ON
SENS:AVER:COUN:AUTO ON
SENS:AVER:COUN:AUTO:TYPE RES
SENS:AVER:COUN:AUTO:RES 3
CALC:MATH "(SENS1)"
CALC:REL:STAT OFF
The command can also be used with a list of parameters, which are separated by a
comma and allow the measurement accuracy and the addressed sensor to be
selected:
MEAS? DEF, <resolution>, <source_list>
The first parameter should always assume the DEF value; it is ignored and is available
only for reasons of compatibility.