Remote Control - Commands
NRP2
198User Manual 1173.9140.02 ─ 07
Query: *PSC?
The query form *PSC? reads out the contents of the power-on-status-clear flag. The
response can be 0 or 1.
Parameters:
<Function>
0 | 1
0
Causes the contents of the status registers to be preserved.
Thus a service request can be generated when switching on the
instrument, if the status registers ESE and SRE are suitably con-
figured.
1
Resets the registers.
*RCL <value>
ReCaLl calls the device state which has been stored with the *SAV command under
the specified number. The effect of *RCL 0 to 19 is the same as if one of the setups 0
to 19 had been selected in manual control. *RCL 0 resets the device to the default
state.
Parameters:
<Value> 1 to 19
Available recall numbers for device setups.
Usage:  Event
*RST
ReSeT sets the instrument to a defined default status. The command essentially corre-
sponds to command SYSTem:PRESet and to pressing (PRE)SET.
The following table documents the difference. SYSTem:
PRESet acts like the frontpanel
key (PRE)SET.
Table 6-1: Difference between commands *RST und SYSTem:PRESet
INITiate:CONTinuous SENSe:AVERage:TCONtrol
*RST OFF REPeat
SYSTem:PRESet ON MOVing
The following table lists all preset values
Common Commands to IEEE 488.2