Rohde & Schwarz NRP2 - 2.4 Measuring average power (Cont Av mode)

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started
NRP2
29User Manual 1173.9140.02 ─ 07
If zeroing fails, the most frequent cause for this error is a signal that is applied to
the sensor and prevents zeroing.
2.4 Measuring average power (Cont Av mode)
In the following example for measuring the average power the following steps are per-
formed:
Setting up the measurement
Setting the frequency
Setting the unit for the measurement result
Setting a fixed offset correction
Performing a relative power measurement
Checking the settings with full height window
Selecting the display mode
Selecting the indication of auxiliary values
Setting up the measurement
To set up the measurement
1.
Put the R&S
NRP2 into operation as described in chapter 1, "Putting into Opera-
tion", on page 7 and connect a sensor to connector A.
2. Preset the R&S NRP2 and execute Zero A.
After preset, one result window indicates the result (in dBm) obtained with sensor
A.
Fig. 2-3: Result window (dBm)
Setting the frequency
The R&S NRP2 must be set to the carrier frequency of the applied signal if the speci-
fied measurement accuracy is to be reached.
Measuring average power (Cont Av mode)