Remote Control - Commands
NRP2
257User Manual 1173.9140.02 ─ 07
Parameters:
<lower_limit> float_value
Range: -200 DB .. 200 DB
*
RST: 0 DB
Default unit:
DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs
(see figure 6-8 Characteristic of analog outputs). The unit corresponds to the output
unit of the measured value.
Value range: See table 6-10 Value ranges, depending on the unit.
Unit: table 6-11 Output unit of measured value of calculate block.
Parameters:
<upper_limit> float_value
*
RST: 20 DBM or 10 DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power (unit DBM, W or DBUV) as the measured
value (see figure 6-8 Characteristic of analog outputs).
Unit:DBM | W | DBUV
Parameters:
<upper_limit> float_value
Range: -120 DBM to 90 DBM
*RST: 20 DBM
Default unit:
DBM
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power ratio (unit DB, DPCT or O) as the mea-
sured value (see figure 6-8 Characteristic of analog outputs).
Unit:DB | DPCT | O
Parameters:
<upper_limit> float_value
Range: -200 DB .. 200 DB
*RST: 10 DB
Default unit:
DB
OUTPut