Rohde & Schwarz NRP2 - Output:recorder<[1]..2>:Limit:upper <Upper_Limit

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
257User Manual 1173.9140.02 ─ 07
Parameters:
<lower_limit> float_value
Range: -200 DB .. 200 DB
*
RST: 0 DB
Default unit:
DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs
(see figure 6-8 Characteristic of analog outputs). The unit corresponds to the output
unit of the measured value.
Value range: See table 6-10 Value ranges, depending on the unit.
Unit: table 6-11 Output unit of measured value of calculate block.
Parameters:
<upper_limit> float_value
*
RST: 20 DBM or 10 DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power (unit DBM, W or DBUV) as the measured
value (see figure 6-8 Characteristic of analog outputs).
Unit:DBM | W | DBUV
Parameters:
<upper_limit> float_value
Range: -120 DBM to 90 DBM
*RST: 20 DBM
Default unit:
DBM
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio <upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power ratio (unit DB, DPCT or O) as the mea-
sured value (see figure 6-8 Characteristic of analog outputs).
Unit:DB | DPCT | O
Parameters:
<upper_limit> float_value
Range: -200 DB .. 200 DB
*RST: 10 DB
Default unit:
DB