Remote Control - Commands
NRP2
233User Manual 1173.9140.02 ─ 07
Usage: Query only
CALCulate<[1]..8>:PTPeak:DATA?
Queries the peak-to-peak distance (maximum to minimum) of the measured values of
the calculate block.
The unit is the current unit of the measured result.
Possible Units: W | DBM | DBUV | DPCT | DB | O
Usage: Query only
CALCulate<[1]..8>:RELative[:MAGNitude] <value>
Determines a value that is used as a divisor (logarithmic subtraction) for all measured
values of the calculate block in the CALC:REL:STAT
ON state. The unit of this value is
equal to the unit of the output value of the calculate block and is defined according to
the following table:
Table 6-11: Output unit of measured value of calculate block
CALC:MATH? CALC:REL:STAT OFF CALC:REL:STAT ON
"(SENSn)"
DBM | W | DBUV DB | DPCT | O
"(SENSn-SENSm)"
DBM | W | DBUV DB | DPCT | O
"(SENSn+SENSm)"
DBM | W | DBUV DB | DPCT | O
"(SENSn/SENSm)"
DB | DPCT | O DB | DPCT | O
"SWR(SENSn,SENSm)"
O -
"REFL(SENSn,SENSm)"
O -
"RLOS(SENSn,SENSm)"
DB -
The value range depends on the current output unit of the calculate block's measured
value, as shown in table 6-10 Value ranges, depending on the unit.
Parameters:
<value> float_value
*RST: 0 DBM or 0 DB
Default unit: See table referred to above.
CALCulate<[1]..8>:RELative[:MAGNitude]:AUTO <auto>
Sets the current measured value as the reference value. When used as a setting com-
mand, this command can only be called by the parameter ONCE. OFF will be ignored.
OFF
is always returned in response to a query.
Error messages:
-230 "Data corrupt or stale": The current measured value is invalid.
CALCulate (Configuration of Calculate Blocks)