Remote Control - Commands
NRP2
272User Manual 1173.9140.02 ─ 07
Like "XPOWer:CCDF", except that a PDF is calculated instead
of a CCDF.
"XTIMe:POWer"
A sequence of measurements is performed. The individual mea-
sured values are determined as in the ContAv mode. The length
of an individual measurement is determined from the ratio of the
total time (
SENS:TRAC:TIME) and the number of individual test
points (SENS:TRAC:POIN).
*RST: depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:APERture <integration_time>
Determines the integration time for a single measurement in the ContAv mode. To
increase the measurement accuracy, this integration is followed by a second averaging
procedure in a window with a selectable number of values. The filter window is config-
ured with the [SENSe<[1]..4>]:AVERage
commands.
Parameters:
<integration_time> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>][:POWer][:AVG]:BUFFer:SIZE <number>
Sets the number of desired values for the buffered ContAv mode.
Parameters:
<number> int_value
Range: 1...400000
*RST: depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:BUFFer:STATe <state>
Switches on the buffered ContAv mode, after which data blocks rather than single
measured values are then returned. In this mode a higher data rate is achieved than in
the non-buffered ContAv mode.
The number of desired measured values is set with the
SENSe<[1]..4>:POWer:
ARR:SIZE command.
Parameters:
<state> ON | OFF
*RST: depending on sensor
SENSe (Sensor Commands)