EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 280

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
277User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>][:POWer]:TSLot:COUNt <number>
Sets the number of simultaneously measured timeslots in the Timeslot mode (see fig-
ure 6-13 Timeslot parameters).
Parameters:
<number> int_value
Range: depending on sensor
*RST: depending on sensor
[SENSe<[1]..4>][:POWer]:TSLot[:EXCLude]:MID:OFFSet[:TIME] <time_interval>
Together with SENS:TSL:MID:
TIME, allows a time interval in the timeslots to be
blanked out. SENS:TSL:MID:OFFS:TIME determines the distance from the start of
the timeslots to the start of the interval to be blanked out (see figure 6-13 Timeslot
parameters).
Parameters:
<time_interval> float_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>][:POWer]:TSLot[:EXCLude]:MID[:STATe] <state>
Activates or deactivates the blanking out of time intervals in the timeslots. There is no
direct command equivalent in the command groups of the power sensors. Instead, the
length of the interval is set to zero.
Parameters:
<state> ON | OFF
Range: depending on sensor
*
RST: depending on sensor. If the *RST value of
SENS:TSL:MID:TIME is zero, the *RST value of
SENS:TSL:MID:STAT is OFF, otherwise it is ON.
[SENSe<[1]..4>][:POWer]:TSLot[:EXCLude]:MID:TIME <time_interval>
Together with SENS:TSL:MID:
OFFS:TIME, allows a time interval in the timeslots to
be blanked out. SENS:TSL:MID:TIME determines the length of the interval to be
blanked out (see figure 6-13 Timeslot parameters).
Parameters:
<time_interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
SENSe (Sensor Commands)