EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 283

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
280User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:SGAMma:CORRection:STATe <state>
Activates/deactivates the measured-value correction of the reflection coefficient effect
of the source gamma.
Parameters:
<state> ON | OFF
*
RST: depending on sensor
Usage: Setting only
[SENSe<[1]..4>]:STATistics:AVERage? <power_value>
Returns the average power value calculated during a statistics measurement (CCDF/
PDF).
Unit: DBM | W | DBUV
Parameters:
<power_value> float_value
Default unit: DBM
Usage: Query only
[SENSe<[1]..4>]:STATistics[:EXCLude]:MID:OFFSet[:TIME] <time_interval>
Together with SENS:
STAT:MID:
TIME, allows a time interval within the time window to
be blanked out; in this interval, the data for the statistics is collected.
SENS:STAT:MID:OFFS determines the distance from the start of the measurement to
the start of the interval to be blanked out (see also [SENSe<[1]..4>][:POWer]:
TGATe<[1]..4>[:EXCLude]:MID:OFFSet[:TIME] on page 274).
Parameters:
<time_interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S
[SENSe<[1]..4>]:STATistics[:EXCLude]:MID:TIME <time_interval>
Together with SENS:STAT:MID:
OFFS, allows a time interval within the time window to
be blanked out; in this interval, the data for the statistics is collected.
SENS:STAT:MID:TIME determines the length of the interval to be blanked out (see
also [SENSe<[1]..4>][:POWer]:TGATe<[1]..4>[:EXCLude]:MID:TIME
on page 275).
Parameters:
<time_interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S