Remote Control - Commands
NRP2
306User Manual 1173.9140.02 ─ 07
Bit No. Meaning
6 Sensor B Front/Rear
Indicates whether sensor B is connected at the front (bit is not set) or rear (bit is set) of the
device.
7 Sensor C connected
8 Sensor D connected
9 Sensor C error
Sensor C is erroneous.
10 Sensor D error
Sensor D is erroneous.
11 Sensor C Front/Rear
Indicates whether sensor C is connected at the front (bit is not set) or rear (bit is set) of the
device.
12 Sensor D Front/Rear
Indicates whether sensor D is connected at the front (bit is not set) or rear (bit is set) of the
device.
13 Not used
14 Key pressed
This bit is always set if a key on the front panel is pressed (CONDition) or was pressed
(EVENt).
15 Bit 15 will never be used.
6.13.8 Questionable Status Register
This register contains information on questionable device states. Such states may
occur when the device is not operated in compliance with its specifications. The regis-
ter can be read by the queries STATus:QUEStionable:
CONDition? or
STATus:QUEStionable[:EVENt]?.
Table 6-20: Meaning of bits used in the Questionable Status Register
Bit No. Meaning
0 to 2 Not used
3 Questionable Power Status Register summary bit
Corresponds to the summary bit of the Questionable Power Status Register.
4 Questionable Window Status Register summary bit
Corresponds to the summary bit of the Questionable Windows Status Register.
5 to 7 Not used
8 Questionable Calibration Status Register summary bit
Corresponds to the summary bit of the Questionable Calibration Status Register.