Table 1-8: GSM/EDGE
Sensor parameter Remote control command in
short form
Value
Sensor Mode
SENS:FUNC POW:XTIM
Aperture Time
SENS:APER
4.615 ms
Duty Cycle
SENS:DCYC
11.762 %
Duty Cycle Correction
SENS:DCYC:STAT ON
Smoothing
SENS:SMO OFF
Dropout Tolerance
SENS:BURST:DTOL
577 µs
Exclude Time Start
SENS:TIM:EXCL:STAR
20 µs
Exclude Time End
SENS:TIM:EXCL:STOP
30 µs
Number Of Timeslots
SENS:TSL:COUN
8
Timeslot Width
SENS:TSL:WIDT
577 µs
Timeslot Fence
SENS:TSL:MID OFF
Timegate Offset
SENS:TGAT:OFFS:TIME
1: 20 µs
2: 2.328 ms
Timegate Length
SENS:TGAT:TIME
1: 527 µs
2: 527 µs
Timegate Fence
SENS:TGAT:MID OFF
Trace Offset Time
SENS:TRAC:OFFS:TIME
-50 µs
Trace Capture Time
SENS:TRAC:TIME
5 ms
Trace Number Of Points
SENS:TRAC:POIN
261
Trace Realtime
SENS:TRAC:REAL OFF
Trigger Delay
TRIG:DEL
0 s
Trigger Level
TRIG:LEV
-20 dBm
Trigger Source
TRIG:SOUR INT
Trigger Slope
TRIG:SLOP POS
Trigger Holdoff
TRIG:HOLD
27 ms
Trigger Dropout Time
TRIG:DTIM
0 s
Trigger Hysteresis
TRIG:HYST
3 dB
Standard Configurations