Getting Started
NRP2
38User Manual 1173.9140.02 ─ 07
Fig. 2-19: Result window / D&A / full height
4. Now, close window 1 with the "Close" softkey.
The "Window" menu shows that window "1" is still selected.
The labeling next to the third softkey is now "Open | Init".
5. Press the "Open" softkey.
The appearance of the window has not changed, and all window-specific settings
have been maintained.
6. Close window
"1" again with "Close".
7. Open it this time with "Init"
The window again displays the measurement function A in digital format.
In contrast to "Open", where the settings remain unchanged, "Init" sets all the
parameters to their default values. "Init" only resets the window settings made in
the "Windows" and "Measurement" menus, but it does not affect the sensor set-
tings!
Window handling