4.2.1.1 Mode dialog box
► Select "Mode..." in the "Sensor" menu.
Fig. 4-4: Mode dialog box
In addition to the "Cont Av" mode (continuous average-power measurements) availa-
ble for all sensors, further measurement modes can be used for most of the sensor
types.
A | B | C | D
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:..., i.e. choose suffix 1 for channel A, channel 2 for B, and so on.
ContAv, Trace & …
Measurement modes:
"Cont Av"
In the Continuous Average mode, the average signal power is con-
tinuously measured without the acquisition interval and signal being
in sync. (see chapter 4.2.1.2, "Continuous average-power measure-
ments", on page 66).
"Trace&Statis-
tics"
In the
Trace & Statistics mode, the R&S NRP2 analyzes the power
envelope of the test signal and displays the power envelope either as
a function of time or in its probability distribution and probability den-
sity distribution function (see chapter 4.2.1.6, "Trace & statistics mea-
surements", on page 72).
"T'slot Av"
In the
Timeslot Average mode, the average power can be measured
over defined time intervals of a signal with periodic envelope.
The timeslot mode should be used preferably for signals with an
equally spaced timeslot structure (see chapter 4.2.1.4, "Measuring
the average power of TDMA signals", on page 68).
"T 'gate Av"
In the Timegate Average mode, the average power can be mea-
sured over defined time intervals of a signal with periodic envelope.
The timegate mode provides four user-configurable gates for the
measurement (see chapter 4.2.1.5, "Measuring the average power in
gates", on page 71 ).