Command Reference
R&S
®
ZNB/ZNBT
1131User Manual 1173.9163.02 ─ 53
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency <RefFreq>
Defines the reference frequency for the frequency-dependent part of the offset loss
([SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet).
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<RefFreq> Reference frequency
Range: Frequency range of the analyzer model.
Increment: 1 MHz
*RST: 1000000000 Hz (= 1 GHz)
Default unit: Hz
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Loss at DC / Loss at Freq / Freq for Loss" on page 578
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet <OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<OffsetLoss> Frequency-independent part of the offset loss
Range: -200 dB to +200 dB
Increment: 0.001 dB
*RST: 0 dB
Default unit: dB
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Loss at DC / Loss at Freq / Freq for Loss" on page 578
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe] <Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean> The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
SCPI Command Reference