GUI Reference
R&S
®
ZNB/ZNBT
280User Manual 1173.9163.02 ─ 53
Balanced Ports...
Opens a dialog to define a balanced port configuration.
See Chapter 6.2.1.3, "Balanced Ports Dialog", on page 256.
6.2.10 Power Sensor Tab
Allows you to set up and perform measurements using external power sensors.
The standard test setup for a "Power Sensor" measurement involves one analyzer
source port and a power sensor. The power sensor is connected to the VNA (e.g. to
the analyzer's USB port) and provides scalar wave quantity results. See Chap-
ter 5.7.20, "External Power Meters", on page 237.
Power Meter
Shows a list of all power meters that have been properly configured. See "Configured
Devices" on page 693.
The bordered label below displays the type and serial number of the selected power
meter.
Remote command:
CALCulate<Ch>:PARameter:MEASure 'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine 'TraceName', 'PmtrD1 | ...
Auto Zero
Initiates an automatic zeroing procedure of the power meter which must be disconnec-
ted from the RF power; see "Zeroing" on page 238. A message indicates that zeroing
is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Source Port
Selects one of the available test ports of the analyzer as a source of the stimulus sig-
nal.
Remote command:
CALCulate<Ch>:PARameter:MEASure 'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine 'TraceName', 'PmtrD1 | ...
Meas Softtool