GUI Reference
R&S
®
ZNB/ZNBT
683User Manual 1173.9163.02 ─ 53
Settling Delay / Reset Delay ← Power Reduction at Sweep End
If Power Mode at Sweep End is set to "Reduce" or "Keep", the "Settling Delay" defines
the time between Restart Sweep request and sweep start. See Figure 6-72 for an illus-
tration.
Use the "Reset Delay" button to adjust the "Settling Delay" to its default value.
Remote command:
SOURce:POWer:SWEepend:SDELay
External Generators > Minimum Settling Delay
Defines the minimum settling delay for external generators. The delay is enforced:
●
whenever settings are applied to a driving generator port, i.e. as a minimum delay
between the setting and the actual measurement
●
irrespective of the sweep type, and whether the generator actually sweeps
Factory default is 180 µs.
Remote command:
SOURce:POWer:GENerator:SDELay
Recovery Tab
Provides a single function "Boot into Recovery Screen" that allows you reboot the
instrument and launch the "R&S Recovery Environment". From there you can restore
the original system image (including operating system and firmware).
See Chapter 11.2, "System Recovery", on page 1381.
6.17.1.3 Info Dialog
The "Info" dialog displays information about the instrument and its operation. All func-
tions are primarily intended for error diagnostic and service purposes; see Chap-
ter 10.3, "Collecting Information for Technical Support", on page 1377. Many "Info"
tabs also display softkeys for printing the contents or saving them to a file.
Access: SYSTEM – [SETUP] > "Setup" > "Info..."
Setup Softtool