Command Reference
R&S
®
ZNB/ZNBT
1244User Manual 1173.9163.02 ─ 53
SOURce:POWer:SWEepend:SDELay............................................................................1245
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]...................................... 1245
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:OFFSet............................................ 1246
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:SLOPe............................................. 1247
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet...................1247
SOURce<Ch>:POWer<PhyPt>:CORRection:LEVel:OFFSet............................................. 1248
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:OFFSet............................................... 1249
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:PERManent[:STATe]............................. 1250
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:STATe................................................. 1250
SOURce<Ch>:POWer<PhyPt>:PERManent[:STATe]....................................................... 1251
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe].............................................................1251
SOURce<Ch>:POWer<PhyPt>:STATe............................................................................1252
SOURce<Ch>:POWer<PhyPt>:STARt........................................................................... 1252
SOURce<Ch>:POWer<PhyPt>:STOP............................................................................ 1252
SOURce:POWer:GENerator:SDELay <SettlingDelay>
If external generators are used in the measurement, this defines the minimum settling
delay before the external generators are considered to be ready, i.e. the time between
setting the generator to the requested frequency and power and continuing with the
measurement.
This is a global setting.
Parameters:
<SettlingDelay> Settling delay per sweep point.
Factory default is 10 µs (= 10
-5
s).
Default unit: s
Manual operation: See "External Generators > Minimum Settling Delay"
on page 683
SOURce:POWer:REDuce:SDELay <SettlingDelay>
Deprecated; superseded by SOURce:POWer:SWEepend:SDELay
If power reduction at sweep end is enabled (SOURce<Ch>:POWer<PhyPt>:
REDuce[:STATe] ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay> Settling delay
Default unit: s
Example: See SOURce:POWer:SWEepend:MODE on page 1244
SOURce:POWer:SWEepend:MODE <SweepEndModes>
Selects the power mode at sweep end.
SCPI Command Reference