Command Reference
R&S
®
ZNB/ZNBT
1243User Manual 1173.9163.02 ─ 53
Setting parameters:
<Trace> String parameter for the trace name, e.g. 'Trc1'. The trace must
exist in the active recall set.
*RST: n/a
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
Usage: Setting only
Manual operation: See "Get Trace..." on page 497
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:INSert<ListNo>
<Frequency>, <TransCoeff>
Inserts a new frequency and transmission coefficient at position no. <ListNo> in the
power loss list. The following points are shifted down in the list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo> Number of point to insert in the list.
Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
<TransCoeff> Transmission coefficient
Range: -300 dB to +200 dB
*RST: n/a
Default unit: dB
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 496
SOURce:POWer... (Contd.)
The SOURce:POWer... commands define the power of the internal signal source.
SOURce:POWer:GENerator:SDELay.............................................................................1244
SOURce:POWer:REDuce:SDELay................................................................................ 1244
SOURce:POWer:SWEepend:MODE.............................................................................. 1244
SCPI Command Reference