Command Reference
R&S
®
ZNB/ZNBT
779User Manual 1173.9163.02 ─ 53
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<VerticalOffset> Vertical offset relative to the 0 V level, i.e. the Voltage level of
the center.
Default unit: V
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Mask Center" on page 635
CALCulate<Chn>:EYE:MASK:DATA?
Returns the detailed results of the mask test in the related eye diagram (i.e. the con-
tents of the corresponding result info field).
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram, or if CALCulate<Chn>:EYE:MASK:STATe is OFF.
Use MMEMory:STORe:EYE:MASK:RESults to save these results to an ASCII file.
Suffix:
<Chn>
.
Channel number used to identify the active trace