GUI Reference
R&S
®
ZNB/ZNBT
571User Manual 1173.9163.02 ─ 53
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:
EMBedding<LogPt>[:STATe]
File Name 1
The ellipsis button in the "File Name 1" column is enabled if a 2-port data Network is
selected.
When loading the touchstone file (*.s2p), the analyzer by default assumes odd ports
left (VNA side), even ports right (DUT side). However, it is also possible to "Swap
Gates", instructing the analyzer to reinterpret the loaded S-parameters (e.g. S
12
-->
S
21
).
Note: The loaded S-parameter traces are stored in the active recall set. Persisted
recall sets contain the full embedding and deembedding data so that they can be trans-
ferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
6.13.2 Offset Tab
Defines length offset parameters for each port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see Chapter 6.13.1.2, "Offset Panel", on page 563).
The marker function Zero Delay at Marker function overwrites the offset parameters.
Background information
Refer to the following sections.
Chapter 5.6, "Offset Parameters and De-/Embedding", on page 183
Chapter 5.6.1.1, "Definition of Offset Parameters ", on page 183
Chapter 5.6.1.3, "Auto Length", on page 184
Chapter 5.6.1.6, "Application and Effect of Offset Parameters ", on page 188
Chapter 5.6.1.7, "Offset Parameters for Balanced Ports", on page 189
Chapter 5.6.1.5, "Fixture Compensation ", on page 187
Offset Embed Softtool

Table of Contents