Command Reference
R&S
®
ZNB/ZNBT
1139User Manual 1173.9163.02 ─ 53
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault on page 1139 to use auto-detection instead of defining "test port --> cal
unit port"-connections explicitely
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine or
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt to define (additional) port assignments manually
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 172 for background infor-
mation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1> Test port number 1
<TestPort2> Test port number 2
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1140
Usage: Setting only
Manual operation: See "Default Port Assignment" on page 424
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
<TestPort1>, <TestPort2>...
Similar logic as [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:
DEFine:DEFault, but initially defines the default port assignments only by their
underlying test port set(s); the connected calibration unit ports are auto-detected at
the start of the corresponding calibration sweep.
The number of created assignments can be queried using the [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:COUNt? command.
See Chapter 5.5.5.4, "Multiple Port Assignments", on page 172 for background infor-
mation.
Suffix:
<Ch>
.
Setting parameters:
<TestPort1>
<TestPort2>
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1140
Usage: Setting only
Manual operation: See "Default Port Assignment" on page 424
SCPI Command Reference