Command Reference
R&S
®
ZNB/ZNBT
1142User Manual 1173.9163.02 ─ 53
Manual operation: See "Port Assignment (manual) " on page 423
Test Port Port Assignment 1 Port Assignment 2
2 Cal Unit Port 1 -
3 Cal Unit Port 2 auto-detected
4 - auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>, <TestPort2>...
Similar logic as [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
DEFine, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are auto-detected at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg> Number of the port assignment
Parameters:
<TestPort1> First test port number
<TestPort2> Second test port number ...
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1140
Manual operation: See "Port Assignment (manual) " on page 423
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with Multiple Port
Assignments, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire)
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
Manual operation: See "Apply/Cancel" on page 427
[SENSe<Chn>:]CORRection:PSTate?
Gets the power calibration label of the active trace in channel <Chn>.
SCPI Command Reference