Concepts and Features
R&S
®
ZNB/ZNBT
118User Manual 1173.9163.02 ─ 53
The receivers and generators can be freely assigned, but without reusing the same
(original) physical port in different (redefined) ports.
Redefining physical ports causes a reset and deletes all switch matrix RF connec-
tions
This can be used to insert external components (e.g. external signal separating devi-
ces, power amplifiers etc.) into the signal path to develop custom measurements, e.g.
to test high-power devices and extend the dynamic range.
Amp
Ext
Gen 1
VNA
21
Dual
Directional Coupler
DUT
Redefined Physical Port 1
a wave: b
1
b wave: b
2
Source: Ext Gen 1
Figure 5-4: Redefined physical port 1 (port 2 not used)
Amp
VNA
32
coupler
DUT
Redefined Physical Port 1
a wave: b
2
b wave: b
3
Source: Port 1
Redefined Physical Port 2
a wave: a
4
b wave: b
4
Source: Port 4
1
4
Figure 5-5: Two redefined ports
High Signal Power
When dealing with external signal amplification, make sure that:
the signals fed to the analyzer are within the allowed range
during calibration the calibration standards meet the requirements in terms of their
power handling capacity
Attenuator pads can be used to adapt the power levels.