Concepts and Features
R&S
®
ZNB/ZNBT
192User Manual 1173.9163.02 ─ 53
The following networks are composed of a serial capacitance C or inductance L (as
seen from the test port), followed by a shunt L or C. They are named
Serial C, Shunt L / Serial L, Shunt C / Serial C, Shunt C /
Serial L, Shunt L.
The following networks are composed of a shunt C or L (as seen from the analyzer
port), followed by a serial C or L. They are named Shunt C, Serial L /
Shunt L, Serial C / Shunt C, Serial C / Shunt L, Serial L.
At the GUI, the "capacitance C<i> in parallel with resistance R<i>" circuit blocks can be
replaced by equivalent "capacitance C<i> in parallel with conductance G<i>" circuit
blocks.
In addition, there is also a Shunt L, Shunt C circuit model available, where the
shunt C is defined as a capacitance C in parallel with a conductance G:
Offset Parameters and De-/Embedding