EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 493

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI Reference
R&S
®
ZNB/ZNBT
493User Manual 1173.9163.02 ─ 53
Remote command:
SOURce:POWer:CORRection:PMETer:ID
Auto Zero
Initiates an automatic zeroing procedure of the power meter; see "Zeroing"
on page 238.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
6.11.3.2 Modify Cal Power Dialog
The "Modify Cal Power" dialog adjusts the target power for the power calibration (cal
power) and defines the target power for the reference receiver calibration. These set-
tings are particularly important for test setups involving external attenuators or amplifi-
ers.
Access: CHANNEL – [CAL] > "Pwr Cal Settings" > "Cal Power..."
Related information
The settings in the "Modify Cal Power" dialog are also used to define port-specific
source powers in arbitrary mode; refer to "Modify Cal Power Dialog" on page 524.
The diagram in the center of the dialog visualizes the settings and results below.
Cal Softtool

Table of Contents

Related product manuals