GUI Reference
R&S
®
ZNB/ZNBT
564User Manual 1173.9163.02 ─ 53
Access: Overview Panel > "Offset"
Figure 6-48: Offset Embed dock widget: Offset panel
The "Offset" panel can also be activated by selecting the Offset Tab or One Way Loss
Tab. Refer to its description for background information, additional parameters and
remote commands.
Active
The checkbox in the "Active" column activates/deactivates the configured length and
loss parameters for the respective Port (i.e. adds/removes them to/from the calculation
flow) without changing the parameter values.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
Reset Offsets
The "Reset Offsets" button resets the length and loss parameters for all ports to their
default values.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe]
6.13.1.3 Single Ended Panel
Allows you to activate or deactivate single ended deembedding/embedding for
selected physical ports. For the "2-Port Data" network type, it is also possible to
change the underlying s2p Touchstone file from the dock widget.