SMIQ RF Level
1125.5555.03 E-82.51
Menu selection: LEVEL - ALC
LEVEL
ALC
UCOR
EMF
STATE
ALC OFF MODE
SEARCH ONCE
LEARN TABLE
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
OFF AUTO ON
SAMPLE&HOLD TABLE
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
ALC-ON
Fig. 2-20 Menu LEVEL - ALC (preset setting)
STATE
OFF Internal level control is deactivated. In this state no AM is possible
.
IEC/IEEE-bus command SOUR:POW:ALC OFF
AUTO Normal state. The internal level control is automatically adjusted to the
operating conditions.
IEC/IEEE-bus command SOUR:POW:ALC AUTO
ON Internal level control is permanently switched on.
IEC/IEEE-bus command SOUR:POW:ALC ON
ALC OFF MODE
SAMPLE&HOLD Level recalibration in the ALC OFF mode after the level or
frequency has been set.
IEC/IEEE-bus command SOUR:POW:ALC:SEAR ON
TABLE In the ALC OFF mode correction values are taken from a
table.
IEC/IEEE-bus command SOUR:POW:ALC:SEAR OFF
SEARCH ONCE
Manual short-time switching on of the level control for level calibration in the
ALC OFF and SAMPLE&HOLD operating mode.
IEC/IEEE-bus command SOUR:POW:ALC:SEAR ONCE
LEARN TABLE
Correction values for the ALC OFF MODE-TABLE function are regenerated
(level control voltage obtained from a table).
IEC/IEEE-bus command SOUR:POW:ALC:TABL?