Sweep SMIQ
1125.5555.03 2.398 E-9
2.26.7 LEVEL Sweep
Menu SWEEP - LEVEL offers access to settings for LEVEL sweep.
Menu selection: SWEEP - LEVEL
Fig. 2-247 Menu SWEEP - LEVEL
START LEVEL
Input value of the starting level.
IEC/IEEE-bus command
:POW:STAR -30dBm
STOP LEVEL
Input value of the stop level.
IEC/IEEE-bus command
:SOUR:POW:STOP -10dBm
CURRENT LEVEL
Indication of the current level.
Operating mode STEP: Input value of the level.
STEP
Input value of the step width.
IEC/IEEE-bus command
:SOUR:SWE:POW:STEP 1dB
DWELL
Input value of the dwell time per step
IEC/IEEE-bus command
:SOUR:SWE:POW:DWEL 15ms
MODE
Selection of the sweep operating mode.
IEC/IEEE-bus commands
:SOUR:POW:MODE SWE;
:SOUR:SWE:POW:MODE AUTO;
:TRIG:SOUR SING
EXECUTE SINGLE SWEEP
Starts a single sweep run. This action to be executed is only
indicated and is only effective if MODE SINGLE is selected.
IEC/IEEE-bus command
:TRIG