SMIQ RF Level
1125.5555.03 E-82.53
Menu selection: LEVEL - UCOR
LEVEL
ALC
UCOR
EMF
FREQ
100. 000 000 0
MHz
LEVEL
- 27.0
dBm
UCOR
+ 1 .9
dB
SELECT LIST... CURRENT: UCOR1
FUNCTION FILL INSERT DELETE EDIT/VIEW
INDEX - FREE 70 - LEN 10 FREQUENCY UCOR1
005 105.000 000 0 MHz + 1.9 dB
006 107.000 000 0 MHz + 1.2 dB
007 108.000 000 0 MHz + 1.3 dB
008 109.000 000 0 MHz + 1.5 dB
009 111.000 000 0 MHz + 1.6 dB
010 112.000 000 0 MHz + 1.9 dB
011 113.000 000 0 MHz + 2.0 dB
012 114.000 000 0 MHz + 2.1 dB
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-22 Menu UCOR - LEVEL-EDIT side
2.5.5 EMF
The signal level can also be set and indicated as the voltage of EMF (open-circuit voltage).
EMF is displayed in the header field of the display after the unit of the level indication.
Menu selection: LEVEL - EMF
LEVEL
ALC
UCOR
EMF
STATE
FREQ
100. 000 000 0
MHz
LEVEL
14.1
mV
OFF ON
emf
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-23 Menu LEVEL-EMF
STATE
ON Voltage value of the level is the voltage of EMF.
OFF Voltage value of the level is voltage at 50 Ω (preset setting).