SMIQ Digital Standard NADC
1125.5555.03 2.275 E-9
(SELECT SLOT...)
TCH
SYNC
Input value for the 28 bit sync word in hexadecimal form.
Depending on the selected slot, the following sync words are
predefined according to the NADC standard:
SLOT 1
A91DE4A
SLOT 2 A9D127A
SLOT 3 C7E3C0C
SLOT 4 342C3F3
SLOT 5 13E23D1
SLOT 6 DC2EC1D
IEC/IEEE-bus :SOUR:NADC:SLOT2:SYNC A91EE4A
SACCH
Selection of data source for SACCH field (Slow Associated
Control Channel).
PN.. PRBS data according to CCITT with periods
between 2
9
-1 and 2
23
-1.
IEEE
:SOUR:NADC:SLOT3:SACC PN15
DLIST Data from a programmable data list.
IEEE
:SOUR:NADC:SLOT3:SACC DLIS
SERDATA Data from data input SER DATA.
IEEE
:SOUR:NADC:SLOT3:SACC SDAT
CDVCC
Input value for the 12 bit "Coded digital verification color
code" field in hexadecimal form. SMIQ does not perform any
coding but directly accepts the input data. 013 hex. is preset.
This corresponds to a value of 1 for the uncoded colour code
DVCC.
IEC/IEEE-bus
:SOUR:NADC:SLOT2:CDVC #HFFF
RSVD
(DOWNLINK only)
Input value for the 12 bit "Reserved" field in hexadecimal
form.
IEC/IEEE-bus
:SOUR:NADC:SLOT2:CDVC #HFFF
SELECT DATA
LIST...
Opens a window for selecting a stored data list or for
generating a new list.
COPY CURRENT
DATA LIST TO...
Stores the current data list under a different name.
DELETE DATA
LIST...
Deletes a data list.
EDIT DATA
LIST...
Opens a window for editing a data list bit-by-bit. The
available storage capacity and the length of the current list is
displayed in parameters FREE and LEN (see also Section
List Editor).
COPY Copies a list range
FILL Fills the range with filler pattern
INSERT Inserts a list range at a different position of the
list
DELETE Deletes a list range
EDIT/VIEW Edits or views the list