SMIQ Digital Standard DECT
1125.5555.03 2.333 E-9
(SELECT SLOT) DATA
Selection of the data source for the 424-bit data field DATA.
The B data field is only displayed for the slot type
ALL_DATA.
PN.. PRBS data according to CCITT V52 or Rec.
0.151 with period lengths between 2
-1 and
2
23
-1.
IEEE-bus
:SOUR:DECT:SLOT3:DATA PN15
DLIST Data from a programmable list.
IEEE-bus
:SOUR:DECT:SLOT3:DATA DLIS
"name"
:SOUR:DECT:SLOT3:DATA DLIS
SERDATA Data fed in via the entry SER DATA.
IEEE-bus
:SOUR:DECT:SLOT3:DATA SDAT
X
This 4-bit data field can not be edited. It contains a CRC
(Cyclic Redundancy Code) generated from the data in the B
data field according to the DECT standard. This data field is
not available for slot type ALL_DATA.
Z
Switches the Z field on and off. This function is available for
the slot types FULL and DOUBLE.
ON The CRC (Cyclic Redundancy Code) generated
for the X field is repeated in the 4-bit data
field Z.
OFF The Z field is not activated. The guard field is
extended by 4 bits.
IEEE-bus command :
SOUR:DECT:SLOT2:ZFI ON
GUARD
This data field can not be edited and contains zero data. At
the beginning of the GUARD field, the power ramp of the
active slot starts falling.
SELECT DATA
LIST...
Opens a window for selecting a stored data list or for
generating a new list.
COPY CURRENT
DATA LIST TO...
Stores the current data list under a different name.
DELETE DATA
LIST...
Deletes a data list.
EDIT DATA
LIST...
Opens a window for editing a data list bit-by-bit. The
available storage capacity and the length of the current list is
displayed in parameters FREE and LEN (see also Section
List Editor).
COPY Copies a list range
FILL Fills the range with filler pattern
INSERT Inserts a list range at a different position of the
list
DELETE Deletes a list range
EDIT/VIEW Edits or views the list