SMIQ Utilities
1125.5555.03 2.421 E-9
Calibration ALC TABLE
For vector modulation or digital modulation, SMIQ is operated with the internal level control switched off.
In mode ALC MODE - SAMPLE & HOLD, the level is recalibrated for each level or frequency setting. In
mode TABLE, the latter does not apply since the necessary input values are taken from a table. This
table can be regenerated using LEARN TABLE
. This is required to attain an optimum level accuracy
for the given operating temperature.
Menu selection: UTILITIES - CALIB - ALC TABLE
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
ALL
VCO SUM
VECTOR MOD
LEV PRESET
OSC REF
LEVEL
ALC TABLE
LEV ATT
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
LEARN TABLE
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-265 Menu UTILITIES - CALIB - ALC TABLE
LEARN TABLE
Triggers the generation of the new correction value for the function LEVEL-
ALC-ALC OFF MODE TABLE.
IEC/IEEE-bus command :SOUR:POW:ALC:TABL?