SMIQ Fitting the Options
1125.5555.03 E-71.11
Fitting the option ½ Plug the module into the appropriate slot for the FSIM2, lock it and
fasten all screws.
½ Open the air inlets at the housing frame by breaking out the safety
glass plate which belongs to the option.
½ The following coaxial connections have to be made:
FSIM2
Module number
1085.XXXX
FSIM2
Module number
1114.XXXX
FSIM1
Cable from from to
W362 X361 X362 X362
W364 X363 X364 X364
W365 X365 X365 X365
W366 X367 X368 X366
W369 X370 X371 X369
Note: Please store the remaining cables. They will be required if
further options will be installed at a later stage or options will
be removed.
½ The included adhesive label ’Option included’ is to be fixed at the
rear panel of the SMIQ.
Calibrating the
Fading Simulator and the ½Warm-up the instrument for 2 hours.
I/Q Modulators ½ Call up menu UTILITIES - CALIB - VECTOR MOD.
Retrofit for 2-channel fading Cables W388 and W391 have to be repositioned for 2-channel fading.
½ Unplug W388 from X368 of FSIM1 and plug to X368 (1085.XXXX)
or X367 (1114.XXXX) of FSIM2.
½ Unplug W391 from X371 of FSIM1 and plug to X371 (1085.XXXX)
or X370 (1114.XXXX) of FSIM2.