MULTI-PORT OPTICAL POWER METER MPM-210H
43 / 59
"Invalid character”
This occurs when unacceptable characters are received for Command or Parameter.
Unacceptable characters: “%”, “&”, “$”, “#”, “~”
"Invalid separator”
This occurs when an unacceptable character is received as a separator between the
Command and the Parameter
Unacceptable characters: “`”, “;”
"Data type error”
This occurs when the Parameter is not an acceptable data type.
"Parameter not allowed”
This occurs when the number of parameters in the corresponding command is more
or less
"Missing parameter”
This occurs when the number of characters in the Parameter is longer than 18.
"Command header error”
This occurs when the number of characters in the Command is longer than 13.
"Undefined Header”
This occurs when an unsupported command is received.
"Setting conflict"
This occurs when one of the following set up commands, other than “STOP” or
“STAT?, was received before measurement using “MEAS” command is completed..
Commands: AVG, LEV, LOGN, AUTO, WAVE, WMOD, WSET, SPE, LOOP, UNIT
"Data out of range"
This occurs when the parameter is outside the acceptable value.
"Program currently running"
This occurs when the mainframe delivers new commands to the module before the
process of delivering commands to the module and receiving responses is
completed.
"Device specific error"
This occurs when the GPIB Address number that you are trying to set exceeds 32.
"Is not Measurement Module"
This occurs when user attempts to deliver a command to a module (slot) that is not
installed.
"Queue overflow"
This occurs
1. when the Queue space which is used to communicate between internal Tasks is
full of messages and there is no space to store any information.
2. when there is no Queue space to store generated Error information
"Queue empty"
This occurs when there is no following message in the Queue space which is used