MPM-210H OPERATION MANUAL
42 / 59
34. LOGN?
Check measurement data point.
Refer to the 5.6.3 Measurement data setting (LOGN).
35. LOGG?
Read out the logging data.
This command is not available for RS-232 communication.
LOGG? <wsp><value1>,<value2>
<value1>: module number
<value2>: port number
Please refer to ‘5.6.4 LOGG?’ for detail information.
LOGG?
Response: Please refer to ‘5.6.4 LOGG?’ for detail
information.
36. ZERO
Before measuring optical power, run Zeroing to delete
electrical DC offset. Please be careful with incidence light
into Optic Port. When using the current meter module,
the MPM-213, please remove the BNC cable from the
MPM-213. This command action takes about 3 sec so
please run other commands at least 3 sec later.
5.5.5 ERR?
All kinds of errors will be saved into message Que. By using a command ‘ERR?’,
users can pull out error information and read them all.