Step Key (or instruction) Display/Printout
The sample is preheated to 50°C.
After 4 minutes, the sample
is heated to 105°C.
11. Display the analysis information
Info soft key
(The moisture loss during phase 1
is displayed; in this case, 2.00%L)
12. Exit ANALYS. INFO display oo soft key
After an additional 6 minutes, the
sample is heated to 200°C.
13. Display the analysis information
Info soft key
(The moisture loss during phase 2
is displayed; in this case: 5.54%L)
14. Exit ANALYS. INFO display oo soft key
After the analysis time has Analys.Time 14.0 min
elapsed, the footer for FinWt + 3.040 g
moisture analysis is printed. --------------------
Res1 + 2.00 %l
Res2 + 5.54 %l
Res3 + 6.30 %l
--------------------
Res + 13.84 %l
--------------------
13. Display the analysis information
Info soft key
after the analysis time has elapsed
(The moisture loss of phases
1, 2 and 3 and the final result
are displayed)
47