58
Mobile printer PW208NX/PW208mNX
Starting Up the Printer
Gap Sensor (PW208NX only)
Printing > Advanced > Calibrate > Gap Sensor
Adjust the Gap sensor level manually.
The adjustment reference values are as follows:
Low : 0.5 V or less
High-Low : 1.0 V or more
The Gap sensor has the following setting items:
1 Emission Level
2 Receiving Level
3 Slice Level
■ Manual Adjustment of the Gap Sensor
1. Remove the label, and then set the liner. Align the liner so that the media sensor
does not detect the I-Mark (black mark).
2. Close the cover.
3. Select Calibrate > Gap Sensor > Emission Level.
4. Press the
button to change the Emit value so that the Sensor value is
adjusted to 0.5 or lower.
5. Press the
button to decide the value.
6. Select Calibrate > Gap Sensor > Receiving Level.
7. Press the
button to change the Recv value so that the Sensor value is
adjusted to meet the following:
Sensor High value - Sensor Low value
1.0 V
Sensor High value: Label + Liner parts
Sensor Low value: Gap part
Adjust the values so that the Sensor value in the Liner part is approximately 0.2 V
to 0.3 V. Then, pull the media to check that the Label + Liner parts value - the
Gap part value
1.0 V.
8. Press the
button to decide the value.
Emission Level (PW208NX only)
Printing > Advanced > Calibrate > Gap Sensor > Emission Level
Adjust the Gap sensor emission level.
Emit : Initial value : 3
: Setting range : 0 to 7