EasyManua.ls Logo

Schiller AT-10 plus - ECG Amplifier and Patient Cable Integrity Test

Schiller AT-10 plus
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 37
Functional Checks 5
AT-10 plus Service Handbook Functional checks and tests 5.3
Art.-no.: 2.540043 Rev.: a
5.3.8 ECG amplifier and patient cable test
1. With a patient simulator connected, press the ECG key . The electrode
screen is displayed (this screen is also displayed in the service screen (see page
20).
2. Disconnect ECG patient simulator. Check the following:
device beeps 4 times
all lead designations highlighted
the mV reading for all leads is -350 to -550 mV
3. Connect ECG simulator and setup HR to 60 b/min, no arrhytmias.
4. Check the following:
all leads stops blinking
the mV reading for all leads is -15 ...+15 mV
5.3.9 External printer test
Only with the Comms Module. At the time of print this was not available
5.3.10 Communication (RS-232) test
Only with the Comms Module. At the time of print this was not available.
Lead Test (mV)
RA 5
LA 5
C1 5
C2 5
C3 5
C4 5
C5 5
C6 5
C7 497
C8 495
C9 498
C10 498
LOFF 0F00
ECG NIBP Spiro HW- I/O +
Service
Character Colour Printer Key Battery
Softwar
When a standard 10-lead cable is connected, check RA, LA, and C1 to C6 only.
Additionally check C7, C8, and C9 when a 13 lead patient cable is used.
Additionally check C7, C8, C9 and C10, when a 14 lead patient cable is used.
LOFF is a register for factory use only (for lead confirmation).

Table of Contents

Other manuals for Schiller AT-10 plus

Related product manuals