5 Functional Checks
5.3 Functional checks and tests
AT-10 plus
Page 34
Art.-no.: 2.540043 Rev.: a
5.3 Functional checks and tests
5.3.1 External sight control
6-monthly
Check the following :
Mechanical condition of the device:
no cracks or chips in the casing.
mains, patient and all other cable assemblies are in good condition with no
crushing, chafing or cuts, etc.). All plugs and sockets are straight and in good
condition.
no soiling which could hamper the safety of the device.
12-monthly
Check the following:
Mechanical condition of the device as detailed above.
Voltage selector is set correctly.
Correct fuse rating according table (see page 43).
The following safety labels are on the device and are readable:
Back Panel, type designation and fuse rating label.
Side Panel (patient connector), CF label and ‘attention‘ symbol.
Required equipment
Safety Tester IEC/EN 60601-1
Calibrated ECG Patient Simulator (e.g Müller & Sebastiani MS410 ECG Simulator)
IMPORTANT!
The measurement devices listed above are subject to the instructions according to
ISO 9000 in regards to Test Equipment Control.