Page 53
Unit Settings 8
AT-10 plus Service Handbook ECG settings 8.1
Art.-no.: 2.540043 Rev.: a
Autom. Format
(Automatic mode for
formats 1 and 2)
ECG Printout No Printout No printout of the ECG given at the end of an auto mode
recording (the recording can be stored in the memory
and printed at a later time if required).
4*3 + 1 Rhythm (r1) Leads are printed in a 4 * 3 format at 25mm/s, with the
selected rhythm lead (r1)at the bottom of the page at
25mm/s.
1*12 at 25 mm/s Leads are printed in a 1 * 12 format at 25mm/s.
8*5 + 4*10s Five seconds of 8 Leads, and 10 seconds of four leads
are printed at 25mm/s.
2*6, 25mm/s, 1p Leads are printed in a 2*6 format at 25mm/s (one page).
2*6, 50mm/s, 1p Leads are printed in a 2*6 format at 50mm/s (one page).
2*6, 25mm/s, 2p Leads are printed in a 2*6 format at 25mm/s (two
pages).
2*6, 50mm/s, 2p Leads are printed in a 2*6 format at 50mm/s (two
pages).
Average Cycles No Printout No printout of average cycles.
4*3, 25mm/s + 2 RhyLeads are averaged over the entire 10 second recording
and printed in 4 groups of 3 leads at 25mm/s, with the
two selected rhythm leads (R1, R2) at the bottom of the
page at 25mm/s.
4*3, 50mm/s + 2 RhyFour groups of 3 leads at 50mm/s, with the two selected
rhythm leads (R1, R2) at the bottom of the page at
50mm/s.
2*6, 50mm/s + 2 Rhy Two groups of 6 leads at 50mm/s, with the two selected
rhythm leads (R1, R2) at the bottom of the page at
50mm/s.
Rhythm Lead R1 Select lead for the first rhythm lead on the screen and printout.
Rhythm Lead R2 Select lead for the second rhythm lead on the screen and printout.
Measurements Select yes or no to print a detailed table of measurement results.
Markings Select yes or no to print reference markings on the ECG average cycle print.
A vertical marker shows the beginning and end of P wave and QRS, and the
end of the T wave.
Interpretation Select yes or no to print interpretation statement (C version only).
Programmable
Leads
Lead 1 to Lead 12 In addition to the conventional lead order (Standard or Cabrera), you can
define an individual lead sequence (user defined).
The screen lead display (and printout) is changed in the Lead tab (next tab) .
Also see user guide.
Lead Lead Sequence This sets the lead sequence for manual and auto printouts as well as on the
screen. Set to Standard, Cabrera. or User Defined.
Signals Sequential of Simultaneous.
Auto Centring Yes or No. Printer trace alignment affects both the manual as well as the
automatic printout. A change of the current setting is only valid on the printout
after a new number of printer channels is selected by pressing the key during
or before printing, or after a new lead group is selected by pressing the lead
keys during printing.
Auto Sensitivity (Reduction) Yes or No. In auto mode a default recording sensitivity of 10 mm/mV is set.
However, if the amplitudes are high meaning that the QRS peaks would
overlap, the sensitivity is reduced automatically to 5 mm/mV. When this is set
an ‘A‘ appears in the information line on the bottom of the screen see page 15.
Rhythm Lead Group Print/ don‘t print (On or Off)
Parameter Options Description