1. Precautions during maintenance work
Replacing CPU-A
Restore the backed up E2PROM contents.
Upgrade the Flash ROM, if required.
Replacing RF PS Assy
Sensitive to static electricity. Take care.
Conduct tuning.
Adjust according to the RF power supply adjustment.
Replacing a turbo molecular pump (TMP)
Reset the TMP operating time.
Replacing a rotary pump (RP)
Reset the RP operating time.
Reference
Approximate Maintenance Plan (Units: months)
Item 3 6 9 12 15 18 21 24 27 30 33 36
RP oil * * * * * * * * * * * *
Clean ion-source * * * * * *
GC trap * * * * * *
GC molecular
sieve filter
* * *
RP overhaul * *
Clean lenses * *
Clean pre-rod * *
Detector *
IG *
PG *
TMP overhaul *