ZEBV-5020C
3.
Detailed explanation of CI and NCI
Model
There are three types of QP2010 dependent on the types of ionization.
Ver. 2.1 of the GCMSsolution is required to use the CI, NCI, and DI option.
Description Notes Part Number Remarks
GCMS-QP2010,
EI without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together.
225-10040-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMS-QP2010c,
EI and CI without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together. Ver. 2.1 of the
GCMSsolution is required to use the
CI.
225-10041-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMS-QP2010nc,
EI, CI, and NCI without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together. Ver. 2.1 of the
GCMSsolution is required to use the
CI and NCI.
225-10042-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMSsolution Ver. 2.1
system package
Order this system package together
with GCMS-QP2010 main frame.
Ver. 2.1 of the GCMSsolution is
required to use the CI, NCI, and DI
option.
225-06628-92 C7
Major different points from 5050A
EI and CI (SEI, SCI) can be simulated using the NCI ion source (for the NCI equipped model
only)
Simulated CI (SCI) mode in NCI ion source is added.
Ionization mode can be selected from SEI, SCI and NCI by indication from PC in NCI ion
source.
Note : SEI in CI ion source is not available.
Selection of reagent gases
QP2010 has two inlet lines for reagent gas. Selection of reagent gas is available from two
types of gas by switching the inlet line from PC.
RP2 is unnecessary for CI and NCI
The rotary pump, which is a backing pump for TMPs, evacuates the reagent gas line at the
same time.
Specification
[1] CI
1. Sensitivity 100pg benzophenone
Quasi-molecular ion S/N>150 at M/Z183 (RMS)
Reagent gas: methane, Scan range: 100-250u, Interval: 0.6sec
2. Reagent gas Methane, isobutane, ammonia
3. Ion source Ion source box exchange type
4. Reagent gas
introduction system
2 line system
5. Auto-tuning Capable
6. Max. column flow
rate (He)
5mL/min (restricted compared to EI to allow introduction of the
reagent gas)

Table of Contents