Degree of residue
After driving the triazine out of the column for one hour at 250
M
C, SCAN data
measurements (approx. 1ppm) can be conducted without hindrance.
Aging for 2 to 3 hours at 250
M
C is recommended for microanalysis (SIM
measurement 100ppb or below) but aging overnight is recommended for
ultramicroanalysis (10ppb or below).
Reference)
High-mass-number standard sample TRIS 0.5g
Tris(perfluoroheptyl)-S-triazine-Maas-Spec
P/N 225-09493-01
Standard sample PFTBA 5g Perfluorotributylamine
P/N 225-09493-03