11. S/N Measurements CI MODEOPTION
Measure S/N.(CI MODE)
Benzophenone 100pg/ul (acetone solvent) 1ul
INJ 250SPLITLESS MODE (Sampling Time 2.0 min)
COL: 80℃(2min)−30℃/min260℃(5min
IF: 250IS200℃
He:
during injection and analysis 100Kpa
Tota l f l o w 20ml/min
MS: SCAN MODE mass range M/Z 100250
interval 0.5sec
 detector voltage relative value +0.4 absolute 1.41.6 kV
 measurement time6.011.0 min(solvent elution time 5.5min
Attention
Please confirm that the sensitivity of EI is sufficient, in the case that the sensitivity of CI and
NCI is confirmed.