Chapter 7 Operating Instructions
DME5000
94 © SICK AG • Germany • Subject to change without notice 8009813/ZJV1/2017-05
Menu Structure
Service
Collective output for freely adjustable status messages in "OR" links that enable
conclusions about the device condition.
Active Inactive
Laser
On
(default)
Pre-failure message
Laser diode
Laser OK
Off –
Level
On
(default)
Pre-failure message
Contamination
Not contaminated
Off –
UpperTemp
On
(default)
Overtemperature
monitoring:
Internal device
temperature > upper
set temperature limit
Internal device
temperature < upper
set temperature limit
Off –
LowerTemp
On
(default)
Undertemperature
monitoring:
Internal device
temperature < lower set
temperature limit
Internal device
temperature > lower set
temperature limit
Off –
Plausib(ility)
On Defective measured
value; causes:
– Light beam
interruption
– Contamination
– Optical/electrical
interferences
Measured value OK
Off –
Ready
On
(default)
Initialization,
Hardware error,
Laser off
Ready, measured value
OK
Off –
Bus Status
On Bus error
SSI: no clock signal
PB/DN: no telegram
transfer
RS-422/Hiperface: ‒
Data transfer OK
RS-422/Hiperface:
independent of bus
status
Off
(default)
–

Table of Contents