38 GM32 · Technical information · 8011923 · V 1.4 · © SICK AG
Operation (Specialist Menus)
Subject to change without notice
4.2.3.7 Analog outputs
Menu: GM32/Parameter/Analog output / Display
Parameter settings for the analog outputs are also used to scale the measured value dis-
play (on the operator panel and in SOPAS ET).
● Electrical zero point “Live Zero”: 0 or 4 mA
● Output of control values after check cycle. Checkmark: Yes
Zero and control values are output after the check cycle:
– First of all the zero value for 90 seconds
– Then the control value for 90 seconds.
● Hysteresis: Threshold for switching from measuring range 1 to measuring range 2, or
vice-versa.
● Source: The signal source is defined during the factory calibration.
● Physical unit
● Start/End: Calibrated measuring range 1 or measuring range 2.
● Signal: Digital output on which measuring range 2 is signaled.
● Control value output can be activated or deactivated individually for each of the 16 ana-
log outputs.
Control value: Checkmark means: Output the control value for this channel.
4.2.3.8 Analog inputs
Menu: GM32/Parameter/Analog input
● Signal sources for measured variables temperature T / pressure P)
● Physical unit of the measured variable
● Electrical zero point “Live Zero”: 0 or 4 mA
● Analog input range for signal source: Start / End