Maintenance 101
5 Maintenance
Maintenance schedule
Perform maintenance procedures at the recommended frequency to maintain the
operating efficiency of your analyzer. Procedures marked with an * may require more
frequent performance (described in each procedure).
Before you begin any ISE maintenance task, do the following:
1. At the Menu Panel, select
Maint., then select ISE Operation.
2. Next to Period.wash, select
OFF, then select Set.
After finishing the ISE maintenance tasks, turn the ISE on:
1. Next to Period.wash, select
ON, then select Set.
Refer to the ISE maintenance schedule for recommended frequencies for ISE procedures
marked with an *.
WARNING
Only Siemens-trained users are to
perform maintenance procedures.
CAUTION
To avoid reporting faulty data,
deterioration of reproducibility, or
damage to the ISE module and its
parts, follow the precautions listed
below:
Immediately clean up any spills
or leaks. Repair the source of
leaks.
Do not leave the chloride
electrode wet pack open. The
electrode dries and becomes
inactive.
Always condition electrodes
before first use.
Use only the cleaning agents
named in the cleaning procedures.
Using any other materials can
cause damage to or deterioration of
analyzer parts. After using any
cleaning agent, always clean parts
with DI water.
BIOHAZARD
Wear personal protective
equipment.