Operating the System 65
2 Operating the System
Start the System
Logging on
There are 4 access levels to the system software:
Level Password Comment
user no Intended for routine operation
supervisor yes Access to some system setup and advanced
maintenance features
tech_manager yes Access to most system setup and advanced
maintenance features
You have to contact your technical service provider
or distributor in order to change the default settings
on these windows.
service yes Access for Siemens service personnel
You are automatically logged on at the user level during startup. No operator is required.
If you log on at the supervisor or tech_manager levels, when done, you should log on to
the user level (no password required).
Starting the system
1. After the power is applied and the Windows operating system is loaded, the ADVIA
Chemistry system Startup window displays.
2. Turn on the optional universal rack handler:
a. For the rack handler, set the control panel Standby/On switch to | (ON).
The ready indicator is green.
b. For the universal rack handler, set the display panel Ready/Standby switch to
READY.
The power indicator is green.
3. At the analyzer power panel, set the Operate/Standby switch (
1) to Operate.